LCD Medical Infusion Pumps With Rate Time Drip And Body Weight Mode
Product description
LCD Medical Infusion Pump With Rate Time Drip And Body Weight Mode
Product advantages
Technical parameters
Selectable modes | Rate mode, Body weight mode,Time mode and Drip mode |
Infusion increment | 0.1ml/h |
Injection accuracy | ±3% |
Applicable infusion sets | All brands standard infusion sets |
KVO rate | Adjustable, 1ml/h~5.0ml/h |
Body weight mode | Drug limit: 0.1~999.9mg Dose:0.1~9999.9mg Volume limit: 0.1~999.9ml Weight: 0.1~300kg (Subject to dosage,unit, weight , liquid and infusion) |
Bolus rate | 100~1000ml/h |
Infusion rate range | 1~2000ml/h |
Time mode | Volume limit: 0.1~9999.9ml |
Time: 1~6000 minutes | |
Preset range | 0.1~9999.9ml |
Rate mode | 1~2000ml/h |
Accumulated volume | 0.1~9999.9ml |
Occlusion value | 39.2kPa~137.2kPa, adjustable, pressure show on the graphic |
Classification | Type BF, class II |
Alarm function | Air,OCCLU, over , out of battery, no AC power supply,drop error, CPU communication error,specification error, low battery, installation error, no operation remind, door open. |
Panel instructions | 1,Infusion rate 2,Battery charging indication 3, Infusion preset volume 4, Accumulated infusion volume 5, Battery capacity 6,AC power supply connection indication 7,Over 8,OCCL. Alarm 9,Air 10,Bed 11,Pressure Value 12,Flow rate , etc. |
Battery | Rechargeable lithium polymer battery 7.4V, 3000mA, working for 4 hours(25ml/h) after being fully charged |
Max power consumption | 25VA |
Power supply | AC100~240V 50/60HZ |
Storage condition | Environment temperature: -20℃~+55℃, air pressure (50~106)kPa,relative humidity≦95% |
Operation condition | Environment temperature: 5℃~+40℃, air pressure (86~106) kPa, relative humidity≦80% |
EMC | Meet the requirements of IEC60601-1-2(2007 version) |
Electrical safety | Meet the requirements of EN60601-1:1990+A1:1993+A2:1995 |
Overall maximum size | 125Length *140 Width*205 Height mm |
Gross weight | less than 2.8kg |