Laboratory Glass Burette Borosilicate Glass Burette PTFE Burette with PTFE Stopcock
Product description
The volume of buret is not completely consistent with its marked volume, so it must be calibrated in analytical work requiring high accuracy. Because glass has the characteristics of thermal expansion and cold contraction, the volume of buret is different at different temperatures. During calibration, a common temperature value must be specified, which is the standard temperature. The international standard temperature of glass containers is 20 ℃, that is, the volume of glass containers is calibrated to the actual volume at 20 ℃ during calibration.
The calibration of the buret is the same as that of the volumetric flask. The water weighing method is used, that is, according to the different densities of pure water at different temperatures, the water quality at different scales of the buret is measured according to V=m/ ρ, Calculate the volume of pure water at a new temperature, which is the real volume of the buret at the new scale.